Q.raxx slimline
Q.raxx offers a series of standard configurations covering a wide range of requirements. Custom solutions, for functions or connection equipment, available on request.

Q.raxx A101-8 slimline
8 universal analog inputs for voltage, current, resistance, potentiometer, Pt100, thermocouple, full and half bridge (1/4-bridge optional), IEPE sensor; 24 bit ADC, sampling rate 100 kHz per channel; signal conditioning
Q.raxx A104-16 slimline
16 analog inputs for thermocouple, 24 bit ADC, sampling rate 100 Hz per channel; signal conditioning
Q.raxx A106-8 slimline
8 analog inputs for strain gauge and inductive bridge, LVDT and RVDT; DC and carrier frequency (/AC) excitation selectable; 24 bit ADC, sampling rate 10 kHz per channel; signal conditioning; 1 analog output and 1 digital I/O each channel
Q.raxx A107-16 slimline
16 universal analog inputs for voltage, current, resistance, potentiometer, Pt100, thermocouple, bridge 24 bit ADC, sampling rate 10 kHz per channel; signal conditioning
Q.raxx A108-16 slimline
16 analog inputs for voltage and current, 24 bit ADC, sampling rate 10 kHz per channel; signal conditioning
Q.raxx D101-16 slimline
12 digital inputs and 4 digital outputs, configurable as counter, frequency (2MHz), PWM or timing input, frequency or PWM output, status input or output; signal conditioning
Q.raxx
The 3U version allows flexible system layout with up to 12 Q.series functions in any desired combination plus Test Controller. For example a 48 multifunctional input at 10 kHz each, data memory, PAC functionality and Ethernet interface in one system.

Q.raxx A101
2 universal analog inputs for voltage, current, resistance, potentiometer, Pt100, thermocouple, full and half bridge (1/4-bridge optional), IEPE sensor; 24 bit ADC, sampling rate 100 kHz per channel; signal conditioning, 2 digital I/O
Q.raxx A102
1 analog input for voltage, current, and half bridges (1/4-bridge optional) and IEPE sensors; 19 bit ADC (SAR), sampling rate 100 kHz; analog output 100 kHz, signal conditioning, 2 digital I/O
Q.raxx A104
8 analog inputs for thermocouple and voltage up to 80 mV, 24 bit ADC, sampling rate 100 Hz per channel; signal conditioning
Q.raxx A106
2 analog inputs for strain gauge and inductive bridge, LVDT and RVDT; DC and carrier frequency (/AC) excitation selectable; 24 bit ADC, sampling rate 10 kHz per channel; signal conditioning, 1 analog output and 1 digital I/O each channel
Q.raxx A107
8 universal analog inputs for voltage, current, resistance, potentiometer, Pt100, thermocouple and bridge; 24 bit ADC, sampling rate 10 kHz per channel; signal conditioning
Q.raxx A108
8 analog inputs for voltage and current; 24 bit ADC, sampling rate 10 kHz per channel; signal conditioning; 2 digital I/O
Q.raxx A109
4 analog outputs, voltage or current selectable, up to 100 kHz, 4 digital in and 4 digital outputs for frequency, counters, incremental sensor, PWM signas, duration or status signal; signal conditioning
Q.raxx A123
4 galvanic isolated inputs for voltage 100 mV, 1 V, 10 V; 24 bit ADC, measuring rate 10 kHz each channel, isolation voltage 1200 VDC; signal conditioning
Q.raxx A124
4 galvanic isolated inputs for thermocouple ; 24 bit ADC, measuring rate 100 Hz each channel, isolation voltage 1200 VDC; signal conditioning
Q.raxx A128
4 galvanic isolated inputs for voltage 40 V, 120 V, 400 V, 1200 V; 24 bit ADC, measuring rate up to 100 Hz each channel by using 2 channels, 50 kHz each channel by using 4 channels; signal conditioning
Q.raxx D101
8 digital inputs and 8 digital outputs, configurable as counter, frequency (2 MHz), PWM or timing input, frequency or PWM output, status input or output; signal conditioning